Specifications
Print

Specifications

Instrument location: neutron-guide hall, thermal beam TG1

Monochromator

  • variable wavelength between 1 - 2.4 x 10-10m ,
  • also ~ 5 Å Δd/d > ~ 2 x 10-3 Collimators

Sample area

  • smallest sample weight ~10mg
  • typical sample size 10mm3 (restricted due to focusing elements)
  • 1s acquisition for 10mm3 (17min for 1mm3) in one shot irreversible experiments
  • 30µs acquisition in stroboscopic mode (reversible experiments)

Detector

Continuous detector bank of 160° Sample environment available

Sample environments available

Ancillary

Performance

Quantity

Cryofurnace-pooled for general use

6K - 800K

3

Cryostat


1.8K-300K

1

Controlled Gas Furnace

100C-1100C

1

Vacuum Furnace

100C-1800C

1

BST uniaxial rig

200kN

1

Vertical Field Cryomagnet / CCR

B= 7.4T T=10K-300K

1

Gas Mixer / Controller

 

1

Sample Changer-Powders

 

1

Eulerian Cradles

 

2